Engineering reference
Viscosity cup
conversion.
Compare Tu-4, DIN 4 and Zahn #2 / #3 cup readings. Add measured density to estimate mPa·s at the same temperature.
How do I compare cup seconds and mPa·s?
Choose the measured cup, enter its flow time and record the sample temperature. Cup equations estimate kinematic viscosity in cSt; mPa·s also needs density measured at that temperature. LK1035 and LK1036 use Tu-4 readings at 25°C. Use these Newtonian approximations for initial screening, then confirm the grade-specific TDS and substrate trial.
Approximate references at 25°C
| Reference | Estimated value | Usable range / condition |
|---|
Cup references describe the same estimated kinematic viscosity at the recorded temperature. They are not measured readings, product acceptance limits or an instruction to dilute a material.
Which Tu-4 readings overlap the international cup ranges?
At 25°C, LK1035 retains its original 20–30 s Tu-4 specification and LK1036 retains 18–28 s. This chart uses only the selected 23–<150 s conversion branch. Their lower readings are not converted here. DIN 4 and Zahn outputs appear only where the estimated result falls inside the respective cup range.
| Tu-4 (s) | Kinematic (cSt) | DIN 4 (s) | Zahn #2 (s) | Zahn #3 (s) | mPa·s only if ρ = 1.00 g/mL |
|---|---|---|---|---|---|
| 23 | ≈76.2 | Outside range | ≈35.8 | Outside range | ≈76.2 |
| 25 | ≈85.2 | Outside range | ≈38.3 | Outside range | ≈85.2 |
| 28 | ≈98.7 | ≈25.5 | ≈42.2 | Outside range | ≈98.7 |
| 30 | ≈107.6 | ≈27.2 | ≈44.7 | Outside range | ≈107.6 |
| 40 | ≈152.5 | ≈36.1 | ≈57.6 | ≈20.5 | ≈152.5 |
| 60 | ≈242.2 | ≈54.8 | Outside range | ≈28.2 | ≈242.2 |
| 100 | ≈421.5 | ≈93.3 | Outside range | ≈43.5 | ≈421.5 |
The final column is a mathematical illustration using 1.00 g/mL, not the density of LK1035, LK1036 or any other grade. For your material, multiply cSt by its measured density in g/mL at the same temperature.
| Grade | Original supplied specification | What this tool can compare |
|---|---|---|
| LK1035 | 20–30 s, Tu-4 cup, 25°C | Only the 23–30 s portion; 20–<23 s is outside the selected branch. |
| LK1036 | 18–28 s, Tu-4 cup, 25°C | Only the 23–28 s portion; 18–<23 s is outside the selected branch. |
| LK1028, LK1022, LK1024, LK1025, LK1033, LK1034, LK1029 | Viscosity is supplied in seconds without a confirmed cup method in the provided specification. | Confirm cup type and temperature in the latest TDS before using this calculator. |
Which equations and limits does this calculator use?
The calculator uses the Biuged Tu-4 relationship from 23 to less than 150 s, DIN 4 from 25 to 150 s, and Elcometer’s Zahn #2 and #3 approximations from 20 to 80 s. These equations estimate cSt. Dynamic viscosity additionally requires measured density at the same temperature.
| Reference | Equation | Allowed input / output |
|---|---|---|
| Tu-4 | ν = (t − 6) / 0.223 | 23 ≤ t < 150 s; selected Biuged branch |
| DIN 4 | ν = 4.57t − 452/t | 25 ≤ t ≤ 150 s; standard cup, not a mini dip cup |
| Zahn #2 | ν = 3.5(t − 14) | 20 ≤ t ≤ 80 s; Elcometer-published approximation |
| Zahn #3 | ν = 11.7(t − 7.5) | 20 ≤ t ≤ 80 s; Elcometer-published approximation |
| Dynamic viscosity | μ = ν × ρ | Density and viscosity at the same temperature |
Here t is flow time in seconds, ν is kinematic viscosity in cSt (mm²/s), ρ is density in g/mL, and μ is dynamic viscosity in mPa·s (cP). An inverse equation produces each cup comparison; the output is then checked against that cup’s range.
Use your cup manufacturer’s calibration and procedure. The Tu-4 manufacturer also publishes a separate low-time branch; this tool deliberately does not apply it. Zahn cup families can use different relationships, and the published linear equations are approximations.
Primary technical references
- Biuged BGD 124: Tu-4 cup equation and range
- Elcometer viscosity dip cups: Zahn equations, English pages 5–6
- VWR / Paragon reference material: DIN 4 equation, page 3; density relationship, page 2
- AMETEK Brookfield: viscosity units and Newtonian limitation
- Anton Paar: viscosity and density at the same temperature
What should I check before using a converted value?
Check the cup type, calibration, flow-time range, sample temperature and density before comparing results. LK1035 and LK1036 explicitly specify Tu-4 at 25°C; seconds alone do not identify a method. For production, keep the original specification and verify current TDS/SDS requirements, Newtonian behavior and acceptance criteria through a documented substrate trial.
Can Tu-4 seconds be converted directly to mPa·s?
A Tu-4 reading first estimates kinematic viscosity in cSt. Dynamic viscosity in mPa·s additionally requires density in g/mL at the same temperature. The selected equation covers 23 to less than 150 seconds and assumes Newtonian behavior. It does not establish a measured rotational viscosity for LK1035 or LK1036.
What part of the LK1035 and LK1036 range can this tool compare?
LK1035 is specified at 20–30 seconds and LK1036 at 18–28 seconds, both in a Tu-4 cup at 25°C. This tool covers only 23–30 seconds for LK1035 and 23–28 seconds for LK1036. Lower readings remain valid original specifications but fall outside this tool’s selected conversion branch.
Are Tu-4, DIN 4 and Zahn #2 or #3 interchangeable?
No. Each cup has its own geometry, equation and usable flow-time range. This calculator compares estimated kinematic viscosity, then checks the target cup range. A missing output means that reference is outside its range. The Zahn estimates use the relationships published by Elcometer, not brand-specific EZ or S90 equations.
Does selecting 25°C correct a reading taken at another temperature?
No. The temperature field records the actual measurement condition; it does not apply a temperature correction. Measure the sample and its density at the same temperature. LK1035 and LK1036 specifications use 25°C. Comparing a reading taken at 20°C requires a new measurement at the specified condition or validated material-specific data.
Can this calculator replace the viscosity method in a TDS?
No. Use the latest grade-specific TDS and agreed test method for acceptance. For shear-dependent materials, record the rotational instrument, spindle, speed and temperature when those measurements are required. A cup estimate is an initial screening reference, not a substitute for supplier limits, current SDS review or a substrate trial.
Product values are typical ranges. Production requires the latest grade-specific TDS/SDS and substrate trials. A calculated estimate is not a verified property of a LeadChem grade.